3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 38
3. The Calibrator will now change to the 10 Ω 2 Wire range.
4. Connect the calibrator Voltage (2 Wire output) terminals to the precision
multimeter. To ensure an accurate measurement connect as a 4 wire measurement,
with both positive leads connected together. Ensure that the multimeter is nulled as a
system, including the leads.
5. Press and measure the output.
6. To adjust the output enter the value as measured on the multimeter
E.g. measured output = 10.00045 Ω, use the function control keys
5. The output can also be adjusted by moving the ‘cursor’ to the required digit and
adjusting the output using the up / down arrows or the digital control
OUTPUT
ON
1
0
•
0
0
0
4
ENTER
5