3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 58
PRT
To calibrate the Capacitance parameter, complete the procedure as follows:
NOTE : The PRT output from the 3000A series calibrator is a passive output. This
means that the value on the display is the Temperature (Based on R0=100) value
generated at the terminals.
The Temperature value cannot be ‘altered’ to a different output, the value stored is
simply the value of the resistor and the connections to the terminal.
1.
Select
CALIBRATE
using the soft key
2. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required parameter
‘PRT’ and then press
SELECT
soft key
3. Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required range e.g.
’PRT 100 °C ’ and then press
SELECT
soft key