3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 21
17.
The ‘2V DC’, ’20V DC’, ‘200V DC’ and ‘1kV DC’ ranges are calibrated in the
same manner.
NOTE : The ‘Zero Calibration’ points for both the 200V DC and 1kV DC Ranges
do not occur at 0V. They take place at 5V and 50V respectively.