3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 20
12.
Measure the output, and adjust as required using the process described in
steps
6 - 10
13.
Use the ‘range up’ and ‘range down’ keys to change the calibration point to
‘Negative Full Scale’.
14.
Measure the output, and adjust as required using the process described in
steps
6 - 10
15.
When calibration of this range is complete press the
RANGE
soft key
16.
To continue adjusting other ranges in the DC Voltage function, select ‘DC
Volts’ and select the required range