3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 48
Press the Range Up / Range Down keys until ‘Full Scale Calibration’ is displayed
Measure the output, and adjust as required using the steps previously described.
If the ‘Full Scale Calibration’ has been adjusted, ensure that the ‘Zero Calibration’
point is still in calibration. It may require more than one cycle of adjusting the Zero
and Full Scale calibration points to bring both points into specification.
56. When calibration of this range is complete press the
RANGE
soft key,
Select ‘Active Ω’, and then select the required range.
57. The ’10 Ω’, ‘100 Ω,‘1 kΩ’, ’10 kΩ’, ‘100 kΩ’, ’1 MΩ’ and ‘10 MΩ’ ranges are
calibrated in the same manner.