3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 24
This will remove any changes that have been made to the output of the
calibrator.
11.
Once the output has been adjusted to within specification, the changes can be
stored to long term memory. To store the changes permanently, press the
STORE
soft key. The following 2 screens are displayed briefly to confirm that
the calibration factors have been saved.
After displaying these messages, the shift key will also cease to be illuminated
12.
Using the ‘RANGE UP’ and ‘RANGE DOWN’ keys, change the output to
‘Positive Full Scale’
SHIFT
Note: All the calibration points can be adjusted prior to storing the
calibration factors (
STORE
), however if the calibration routine is
ended or there is a power failure the new calibration factors will not
be saved if
STORE
has not been pressed.