3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 68
3. The Calibrator will now change to the Scope Bandwidth Function. Use the Range
Up / Range Down keys to select the required Scope Bandwidth adjustment point
The currently
selected adjustment point is displayed on the screen above
the soft
key menu
4. Connect oscilloscope to the SCOPE terminal on the front panel. Ensure that the
oscilloscope is set to 50 Ohm input impedance
5. Set the oscilloscope to 100mV/Division
5. Press and measure the output.
6. To adjust the output enter the value as measured on the oscilloscope
E.g. measured output = 610.00 mV, use the function control keys
5. The output can also be adjusted by moving the ‘cursor’ to the required digit and
adjusting the output using the up / down arrows or the digital control
6
1
0
•
0
0
ENTER
OUTPUT
ON