3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 22
D.C. Current
To calibrate the D.C. Current parameter, complete the procedure as follows:
1.
Select
CALIBRATE
using the soft key
2.
Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required parameter
‘DC Amps’ and then press
SELECT
soft key
3.
Use the ‘Digital Control’ or the ‘Arrow Keys’ to highlight the required range e.g.
‘200uA DC’ and then press
SELECT
soft key
4.
Connect the calibrator 2A current terminals to the precision Multimeter. Ensure
that the Multimeter has been nulled with the leads disconnected