3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 52
56. When calibration of this range is complete press the
RANGE
soft key,
Select ‘2 Wire Ω’, and then select the required range.
57. The ‘1nF’, ‘10nF’, ‘20nF’, ‘50nF’, ‘100nF’, ‘10uF’ and ‘100uF’ ranges are
calibrated in the same manner.
The 1mF and 10mF ranges are simulated