3000A SERIES CALIBRATION MANUAL
TRANSMILLE LTD. Version 3 : Jun 2014
Page 79
8. Use the ‘range up’ and ‘range down’ keys to change the calibration point to
‘Zero Calibration’.
9. Measure the output as before; if necessary adjust as previously described above.
11. Measure the output as before; if necessary adjust as previously described above.
56. When calibration of this range is complete press the
RANGE
soft key
Select ‘DC Current’ and then ‘2mA ’ and proceed with this range as described above.
57. The ’2mA DC’, ‘20mA DC’,‘200mA DC’, ‘2A DC’ and ‘30DC’ ranges are
calibrated in the same manner.
Remember to change the connections when calibrating the
30A range