8 Alarm messages, notifications and malfunctions
96
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
Alarm
Pri-
ority
ID
Cause
Remedial action
Technical
error
! !
>100 A technical fault has oc-
curred. The dispensing of
anaesthetic agent is not
possible.
Put the vaporizer into
Standby mode and restart.
Additional information can
be found in chapter 6.3.
Technical
error
! ! !
>100 A technical error has oc-
curred. The dispensing of
anaesthetic agent has
been stopped.
A
Continue sedation of
the patient using an al-
ternative option.
B
Put the vaporizer into
Standby mode and re-
start.
Additional information can
be found in chapter 6.3.
Vaporizer
blocked (inter-
lock)
!
12
The vaporizer is blocked
by the interlock system
from dispensing anaes-
thetic agent.
Unblock the vaporizer.