8 Alarm messages, notifications and malfunctions
94
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
Alarm
Pri-
ority
ID
Cause
Remedial action
Dispensing
DES stopped
(interlock)
! ! !
02
The vaporizer has been
blocked by another vapor-
izer (interlock).
(Only with SmartVap D)
To continue dispensing
anaesthetic agent, un-
block the vaporizer (in-
terlock) and restart the
dispensing of anaes-
thetic agent.
Or press the turning
knob to confirm the
stopping of dispensing
of anaesthetic agent.
Dispensing
DES stopped
(mech. lock)
! ! !
01
The mechanical lock for
the vaporizer was
opened.
To continue dispensing
anaesthetic agent, lock
the vaporizer onto the
connector panel of the
anaesthesia workstation
and restart the dispens-
ing of anaesthetic
agent.
Or press the turning
knob to confirm the
stopping of dispensing
of anaesthetic agent.
Fill level low
!
16
The level in the reservoir
is low.
Fill the vaporizer reservoir.
Fill level low
! !
10
The level in the reservoir
is very low.
Fill the vaporizer reservoir.
Fill opening
open
!
15
The cover flap of the fill
opening has been open
for 5 minutes.
Close the cover flap of the
fill opening.
Fresh gas flow
too high
!
14
The fresh gas flow is
above the specified
range.
Ensure that the fresh gas
flow of the anaesthesia
workstation is within the
specified range of between
0.2 and 15 L/min.
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......