1 Introduction
10
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
1.3.3
Maintenance personnel
Basic maintenance work must always be performed by personnel with electrical/electronic
and mechanical expertise, as well as experience in the maintenance of medical devices.
1.4
Contraindications
Do not use the vaporizer on patients for whom volatile anaesthetics are contraindicated.
For contraindications of Desflurane
, follow the medication’s accompanying documents.
1.5
Place of use
The vaporizer is intended for stationary use in hospitals and rooms used for medical pur-
poses.
WARNING
Risk of fire
Only operate the vaporizer in areas with an oxygen concentra-
tion of < 25 Vol %.
Do not operate the vaporizer in areas where combustible or
explosive gas mixtures could arise.
Do not operate the vaporizer in the presence of flammable an-
aesthetics.
WARNING
Risk to patient from device malfunctions caused by interference
with other electronic medical devices
Do not use the vaporizer in the physical vicinity of an MRI unit.
Do not use the vaporizer in hyperbaric chambers.
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......