7 Menu mode
70
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
7.3
Resetting and downloading use case data
The vaporizer shows the following information in the
Use case data
dialog:
Consumption of Desflurane in mL
Duration of dispensing in h
Average consumption in mL/h
The data can be reset in the
Use case data
dialog. If the vaporizer is put into Standby
mode, or in the event of a complete power failure (no mains power supply and no power
supply from the emergency battery), the data is automatically reset. The data is down-
loaded in the form of a text file.
Prerequisite:
To download the data, a USB drive (FAT32 formatted) with sufficient
memory must be connected.
i
Additional information about the USB interface can be found in chapter 15.
A
Insert the USB drive (1) into the USB port (2) on the vaporizer.
B
Start Menu mode.
► The
Use case data
dialog is displayed.
C
Tap the
Reset
(4) button.
► The
Reset
(4) button will change color to blue.
► The illuminated ring around the turning knob (5) flashes blue.
2
1
5
6
4
3
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......