3 Description
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
25
3.4
SmartVap S - Connection and interlock system
(1)
Locking knob for mechanical locking
(2)
Interlock pins (Selectatec)
(3)
Lock pin for mechanical locking
(4)
USB interface to export consumption data and event log files, and to conduct
SW updates
(5)
Mains power supply connection with cable protection
Interlock system function
Fig.: Right vaporizer (2) in operation, left vaporizer (3) blocked.
SmartVap S (3) is blocked from dispensing Desflurane if the dispensing of anaesthetic
agent is active for an adjacently mounted vaporizer (2) and thus the interlock pins (1) have
been moved out.
1
2
5
4
3
1
1
3
2
Summary of Contents for SmartVap D