8 Alarm messages, notifications and malfunctions
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
95
Alarm
Pri-
ority
ID
Cause
Remedial action
Loss of mains
power
!
17
The vaporizer was discon-
nected from the mains
power supply while dis-
pensing anaesthetic
agent. The charge level of
the emergency battery is
higher than 70%.
If the vaporizer is not
used for dispensing an-
aesthetic agent, it turns
off after a short time.
Connect the vaporizer to
the mains power supply.
Loss of mains
power
! !
09
The vaporizer was discon-
nected from the mains
power supply. The charge
level of the emergency
battery is lower than 70%.
If the vaporizer is not
used for dispensing an-
aesthetic agent, it turns
off after a short time.
Connect the vaporizer to
the mains power supply.
Mechanical
lock open
!
11
The vaporizer is not
locked onto the connector
panel of the anaesthesia
workstation. The vapor-
izer is blocked from dis-
pensing anaesthetic
agent.
Lock vaporizer onto the
connector panel of the an-
aesthesia workstation.
Mechanical
lock open
! !
06
The vaporizer is not
locked onto the connector
panel of the anaesthesia
workstation. A fresh gas
flow has been detected.
The vaporizer is blocked
from dispensing anaes-
thetic agent.
Lock vaporizer onto the
connector panel of the an-
aesthesia workstation.
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......