5 Operation
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
53
A
Unscrew the cap of the anaesthetic bottle (Piramal Fill filling system) and ensure that
the O-ring is intact.
B
Manually open the cover flap (1) of the fill opening on the vaporizer.
C
Screw the open anaesthetic bottle (2) onto the fill opening in a clockwise direction until
the end of the thread is reached.
D
Turn the anaesthetic bottle (2) by another approximately 1/3 turn in a clockwise direc-
tion, and hold.
► The bottle valve is open and the Desflurane flows into the vaporizer reservoir.
E
To end the filling, turn the anaesthetic bottle (2) by an approximately 1/3 turn in a coun-
terclockwise direction.
► The bottle valve is closed.
F
Slowly unscrew and remove the anaesthetic bottle by turning counterclockwise.
CAUTION
If the anaesthetic bottle is unscrewed too quickly, a small amount
of Desflurane may be released into the environment.
1
2
2