5 Operation
56
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
A
Set the dispensing quantity by swiping left on the indicated screen area (1) or by turning
the knob (2) in a clockwise direction.
► The illuminated ring around the turning knob flashes blue.
Setting range for Desflurane dispensing quantity:
0, then 2 to 12, 12 to 18 Vol % in increments of 0.5 Vol %
B
Confirm the set dispensing quantity of Desflurane by briefly pressing the turning knob
(2).
► The illuminated ring around the turning knob shows a circling green point of light.
► The dispensing of Desflurane into the fresh gas flow is initiated.
i
Additional information about setting the dispensing quantity can be found in chapter 4.2.
5.7
Dispensing quantity greater than 12 Vol %
WARNING
Risk to patient from high Desflurane setting values, resulting in
a reduced oxygen concentration in the fresh gas
High Desflurane concentrations can affect the hot wire flow meas-
urement (if present). The expiratory volume displayed is too high in
this case.
Observe the patient’s condition.
1
2
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......