8 Alarm messages, notifications and malfunctions
100
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
8.3
Messages on the operating screen
Message
Description
Dispensing DES not possible.
Place vaporizer in standby mode and re-
start.
In the event of a technical error (alarm
ID 01), this message is displayed as addi-
tional information.
Additional information can be found in
chapters 6.2 and 8.1.
Press the turning knob to enable settings
higher than 12 Vol%.
This message is shown when the dispens-
ing quantity is changed, as soon as the
dispensing quantity is increased above
12.0 Vol%.
Additional information can be found in
chapter 5.7.
Close the mech. lock and restart dispens-
ing DES.
Or press the turning knob to confirm.
This message is shown in addition to an
alarm if the vaporizer was blocked by the
mechanical lock being opened during the
dispensing of anaesthetic agent (alarm ID
01).
Additional information can be found in
chapter 8.1.
Unblock the vaporizer and restart dispens-
ing DES.
Or press the turning knob to confirm.
This message is shown in addition to an
alarm if the vaporizer was blocked by an-
other vaporizer (interlock) during the dis-
pensing of anaesthetic agent (alarm ID
02).
Additional information can be found in
chapter 8.1.
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......