10 Emptying the vaporizer
106
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
10
Emptying the vaporizer
WARNING
Risk of harm to persons and damage to the device from incor-
rect emptying.
The vaporizer must only be emptied by personnel in accord-
ance with chapter 1.3.3.
CAUTION
Risk to persons from contamination of the environment with
volatile anaesthetic.
Avoid spilling liquid volatile anaesthetic.
Never press down manually on the filling valve under the
cover flap of the fill opening.
Ensure proper ventilation of the premises while emptying.
CAUTION
Risk to persons from incorrect handling, storage and disposal
of the volatile anaesthetic.
Emptied anaesthetic agent must be handled, stored, and dis-
posed of as a medication, in accordance with hospital policies
and the statutory regulations of the respective country.
Prerequisite:
The vaporizer is turned off and disconnected from the mains power sup-
ply. The vaporizer is not connected to an anaesthesia workstation.
A
Allow the vaporizer to cool to room temperature.
B
Unscrew the cap of an empty Desflurane anaesthetic bottle (Piramal Fill filling system).
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......