5 Operation
48
SmartVap Vaporizer SW 1.n, Instructions for use (en-US)
E
Connect the power cable to the mains power supply, making sure that the power plug is
accessible at all times.
► The status LED on the front side of the vaporizer will light up green.
►
The vaporizer heats up.
5.3
Connecting to an anaesthesia workstation
WARNING
Risk to patient from overdose or mixing of anaesthetic agents if
multiple vaporizers are connected to an anaesthesia work-
station
Make sure that the interlock system for the vaporizer and an-
aesthesia workstation is operating correctly.
When using the SmartVap S at an anaesthesia workstation
with a connector panel for three vaporizers, the middle posi-
tion on the connector panel must always be occupied by a va-
porizer in order to ensure that the interlock system is opera-
tive.
WARNING
Risk to patient from oxygen deficiency if the vaporizer is incor-
rectly connected to the anaesthesia workstation
Connect the vaporizer as described.
The mechanical lock of a vaporizer mounted on the connector
panel of an anaesthesia workstation must always be in the
locked position.
Observe the instructions for use of the anaesthesia work-
station.
Please note:
Before connecting the vaporizer, make sure that the vaporizer model and
the connector system of the anaesthesia workstation are compatible
(see chapter 3.1).
The illustrations show the connecting of a SmartVap S. The SmartVap D is connected in
the same way.
Summary of Contents for SmartVap D
Page 1: ...Instructions for use SmartVapTM D SmartVapTM S SW 1 n...
Page 2: ......
Page 3: ...SmartVap Vaporizer SW 1 n Instructions for use en US 3 Page intentionally left blank...
Page 126: ...17 PIN information 126 SmartVap Vaporizer SW 1 n Instructions for use en US...
Page 128: ...128 SmartVap Vaporizer SW 1 n Instructions for use en US Notes...
Page 129: ...SmartVap Vaporizer SW 1 n Instructions for use en US 129 Notes...
Page 130: ...130 SmartVap Vaporizer SW 1 n m Instructions for use en US Notes...
Page 131: ......