iSort
™
Automated Cell Sorter User Guide
91
Prepare instrument
for Shut Down
1.
When the Shut Down procedure is completed and the system is powered off,
leave the sample tube of deionized water in the sample tube holder in the up
position.
2.
Remove the nozzle from the holder, then remove the nozzle tip from the
nozzle assembly, and clean it in an ultrasonic bath (sonicator) as described on
page 98.
3.
While the nozzle tip is in the ultrasonic bath, loosely screw the small black nut
back on the nozzle assembly, then cover it with the black rubber protector
containing deionized water.
4.
Remove and clean the deflection plates as described on page 97.