36
iSort
™
Automated Cell Sorter User Guide
5.
To initiate sample loading, move the sample tube holder up until it clicks into
place, close the sorting compartment door, then press
Continue
.
IMPORTANT!
Cell sorting creates aerosols. Exposure to and inhalation of
aerosols can potentially be hazardous depending on the cell types you are
using. Always keep the door to the sorting compartment shut to minimize the
risk, but understand that this does not eliminate the risk entirely.
We recommend placing the iSort
™
Automated Cell Sorter in a biosafety cabinet
when sorting human primary cells and cells known to be infected with viruses
or other agents. Consult your institution’s safety officer for guidelines related
to your individual samples.
6.
When sample loading is initiated, the
Define Region
screen opens and the
fluidic system temporarily increases the sample flow rate to quickly transfer the
cells to the optical detection area.