114
iSort
™
Automated Cell Sorter User Guide
3.
Insert the fittings into the bottom slots in the pump assembly to secure them.
4.
Align the sheath pump tubing to the center of bottom rollers, keeping it away
from the edges where it could possibly get pinched or caught in a gap.
5.
Repeat the process for the top set of the sheath pump tubing, gently sliding it
into position and centering it within the top rollers.
6.
After you have installed both sets of sheath pump tubing and aligned them
into position, pull the clamping bracket towards you to engage the new
tubing.