70
iSort
™
Automated Cell Sorter User Guide
6.
Make sure that the collection tray is pulled out, then press
Continue
.
7.
When prompted, add Attune
™
Performance Tracking Beadst in a sample tube,
then place the sample tube with the calibration beads in the tube holder.
8.
Push the tube holder upward until it stops, then press
Continue
to start the
sample flow.
9.
When prompted, enter the calibration sample volume (200–4500 µL) using the
numeric keypad on the touchscreen, then press
Finished
.
10.
Confirm that the sample volume is entered correctly, then press
Confirm
to
proceed to the next step.