iSort
™
Automated Cell Sorter User Guide
81
9.
After completing the alignment, the instrument automatically measures the
nozzle tilt to ensure that it is within specifications, then measures the gimbal
angle.
10.
If further adjustments are required, the instrument provides you with the
necessary instructions, and rechecks the nozzle tilt for proper alignment.
Follow the instructions provided until the instrument displays the “Gimbal
Adjustment Passed” dialog, then press
Home
on the dialog to return to the
Home screen.