94
iSort
™
Automated Cell Sorter User Guide
9.
Instrument care and maintenance
Planned maintenance schedule
To ensure reliability of the iSort
™
Automated Cell Sorter, you must perform basic preventative maintenance
procedures on a regular basis. In addition, certain maintenance and parts replacement procedures must be
performed by the Thermo Fisher Scientific field service engineers (FSE) annually. The following table lists
the routine maintenance procedures that keep the iSort
™
Automated Cell Sorter in good working condition.
For detailed instructions on how to perform a specific maintenance function, see pages 96–118.
Procedure
Frequency
Materials required
User FSE
Inspect and wipe down bottle
and fluidics compartment area
for drips
Daily
Kimwipes
™
laboratory tissues (or other soft,
lint free cloth)
Perform initial cleaning
Daily and when
prompted
Sheath fluid (1X PBS, flow cytometry grade)
(Cat. No. A1286301), deionized water
Shut down the system
Daily
10% bleach solution, Attune
™
Wash Solution
(Cat. No. A24974), deionized water
Remove and clean the nozzle
tip, store it in cleaning solution
Daily
5% Contrad
™
70 detergent solution (cleaning
solution), deionized water, ultrasonic bath
(sonicator), iSort
™
Nozzle Tip cover (Cat. No.
A33411)
Remove and clean the
deflection plates
Daily
Kimwipes
™
laboratory tissues (or other soft,
lint free cloth), isopropyl alcohol or ethanol
Clean the sort chamber
Between sorts
Kimwipes
™
laboratory tissues (or other soft,
lint free cloth), deionized water
Clean the SIP (sample input
port)
Between sorts
Kimwipes
™
laboratory tissues (or other soft,
lint free cloth)
Perform sample line cleaning
Between sorts
10% bleach solution, Attune
™
Wash Solution
(Cat. No. A24974), deionized water
Perform flow rate calibration
As needed
Attune
™
Performance Tracking Beads (Cat.
No. 4449754), 5% Tween
™
20 in deionized
water, 10X PBS (Cat. No. 70011044), Sheath
fluid (1X PBS, flow cytometry grade) (Cat.
No. A1286301), deionized water
Perform drop delay calibration
As needed
Attune
™
Performance Tracking Beads (Cat.
No. 4449754), 10X PBS (Cat. No. 70011044),
Sheath fluid (1X PBS, flow cytometry grade)
(Cat. No. A1286301), deionized water,
microscope slide (Cat. No. A36699),
fluorescent microscope
Perform gimbal adjustment
As needed
Sheath fluid (1X PBS, flow cytometry grade)
(Cat. No. A1286301)
Summary of Contents for invitrogen iSort Series
Page 120: ...120 iSort Automated Cell Sorter User Guide...
Page 121: ...iSort Automated Cell Sorter User Guide 121...
Page 122: ...122 iSort Automated Cell Sorter User Guide...
Page 123: ...iSort Automated Cell Sorter User Guide 123...
Page 140: ...thermofisher com support thermofisher com askaquestion thermofisher com 19 October 2018...