102
iSort
™
Automated Cell Sorter User Guide
3.
Tighten the black cap with the torque wrench (provided with the instrument)
until you hear two clicks. Do not overtighten.
Note:
Loose coupling of the nozzle tip to the nozzle assembly can result in
decreased system performance.
4.
Reinsert the nozzle assembly into the Nozzle stage. The nozzle assembly
should snap into place by the magnetic fastener.
When the nozzle is placed into its housing correctly, you will hear a click as it
snaps into place. Once you hear the click, you should not be able to rotate the
nozzle.
Note:
To manually adjust the positioning of the nozzle assembly so that the sample
stream is properly aligned with the system optics, see “Gimbal Adjustment”,
page 77.