Insert the sample line through the screw and the clear Lucite block that
secures the sample injection port to the instrument. When the sample line is
positioned correctly, tighten the screw.
4.
Insert the sample pump tubing into the shoe, making sure that the tubing
remains slack.
5.
Insert the sample pump shoe into its slot in the sample pump assembly,
making sure that the sample pump tubing is pressed against the pump rollers.
6.
Tighten the sample pump tubing by pulling the tubing outwards and sliding
the blue and orange stoppers flush against the sample pump shoe.