iSort
™
Automated Cell Sorter User Guide
109
7.
When you are finished replacing the sample pump tubing, close the top the
cover and the sorting compartment door, then power on the instrument.
8.
In the
Maintenance
Tubing Life
screen, select
Sample Pump.
9.
In the Sample Pump Life dialog, press
Replacement Complete
to reset the
sample tube life counter to 100%.
IMPORTANT!
Inspect the new tubing for leaks during the first run after replacement.
Summary of Contents for invitrogen iSort Series