Automated Cell Sorter User Guide
105
2.
Remove the shoe that secures the sample line to the rollers on the back of the
sample pump.
To do this, push the left arm of the shoe inward to release the notch that
secures it to the pump assembly and rotate the shoe away from the pump,
then pull the shoe away from the rollers.
3.
Remove the tubing from the shoe by pulling on the blue and orange stoppers
attached to the tubing line and moving the line away from the shoe.
4.
Loosen the screws that hold the sample pump dampener, but do not yet
detach the dampener from the remaining tubing.