Replace sample pump tubing
In general, we recommend that you replace peristaltic pump tubing for the sample
line (Cat. No. A33303) when prompted by the instrument. Normally, the sample
pump tubing needs to be replaced every 10 weeks based on 20 hours per week
average run time (or every 200 hours of run time).
IMPORTANT!
When handling the tubing, make sure that it does not become
twisted or kinked.
Sample tube dampener
Blue stopper
Orange stopper
Sample line
Sample pump tubing
Remove sample
pump tubing
1.
Shut down the instrument and unplug before replacing any tubing. Once the
Shut Down is complete, open the top cover of the instrument and the sorting
compartment door.
Sample tube dampener
Blue stopper
Screw
Rollers
Sample pump tubing
Orange stopper
Shoe