98
iSort
™
Automated Cell Sorter User Guide
Remove and clean the nozzle tip
Overview
The nozzle is responsible for accepting sample and sheath flows, for producing a
focused and stable stream through the 85-µm nozzle tip, and for creating and
charging droplets.
For optimal performance, the nozzle tip must be removed and cleaned in an
ultrasonic cleaning bath following instrument Shut Down at the end of the day. It is
also possible for the nozzle tip to become clogged during normal operation, which
might require removal and manual cleaning. In such cases, the instrument displays
a warning dialog.
IMPORTANT!
Always use clean gloves when handling the ceramic nozzle tip.
Never handle the nozzle tip with bare hands, because the oils from your skin can
affect fluid flow and instrument performance.
Remove and clean
the nozzle tip
The nozzle assembly is located in the sorting area and is secured to its housing via
magnetic fasteners. Before removing the nozzle assembly, make sure that the
sample and sheath flows have stopped.
Handle
Nozzle assembly
Nozzle stage
1.
Grab the nozzle assembly by the handle and lift it up to remove it from its
housing. Note that it may require some force to overcome the magnetic fastener.