iSort
™
Automated Cell Sorter User Guide
89
2.
Press
Continue
to confirm the Shut Down. The instrument initiates the Shut
Down procedure.
3.
When prompted, place a tube containing 3 mL of 10% bleach solution into the
sample tube holder.
4.
Move the sample tube holder up until it stops, then press
Continue
to start
the Shut Down Cleaning – Step 1.
5.
When prompted, remove the tube containing the 10% bleach solution from the
sample tube holder, and replace with a tube containing Attune
™
Wash Solution.
6.
Move the sample tube holder up until it stops, then press
Continue
to initiate
the Shut Down Cleaning – Step 2.