Unscrew the small black cap that secures the nozzle tip to the nozzle assembly
using the torque wrench provided with the instrument.
3.
Remove the ceramic nozzle tip and the O-ring.
IMPORTANT!
Be very careful when handling the nozzle tip. Hold the nozzle
tip only by its base and never handle it with bare hands. Always use clean
gloves when handling the nozzle tip.
Handle
85-µm nozzle tip
Nozzle assembly
Black cap
O-ring
4.
Place the nozzle tip in a tube containing 5% Contrad
™
70 detergent solution in
deionized water with the nozzle tip facing up.
5.
Cap the tube and place it in an ultrasonic cleaning bath (sonicator) for
30 minutes.