21
Performance Verification
Troubleshooting Checklist
100
GENESYS 10S Vis User Guide
Thermo Scientific
Troubleshooting Checklist
If a Performance Verification test fails, follow the instructions below to diagnose common
problems.
If a test continues to fail after you have tried all these recommendations, follow the
troubleshooting list for the test being run (included with the description of each test).
Make sure:
• You follow the instructions for the test properly.
• Filters and standards are clean.
• The sample compartment door is closed during the test.
• The sample compartment is clear of obstructions.
• The cell holder assembly is installed properly. If the 6-Position Cell Holder is installed,
run the test once with the sample compartment door open to verify that the 6-Position
Cell Holder is moving smoothly.
• No problems are indicated by the power-on diagnostics after you turn the instrument
power OFF and then ON.
• The lamp is ON.
• The lamp compartment is clear of obstructions.
WARNING
Do not open the lamp compartment unless the instrument power is OFF.
WARNING
Do not turn the instrument power ON unless the lamp compartment is
closed.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...