20
Kinetics
Measuring Samples
Thermo Scientific
GENESYS 10S Vis User Guide
93
2. When the parameters are set, press
Save Test
to save the test or
Run Test
to measure a
blank or sample.
Measuring Samples
Y
To measure samples
1. In the Kinetics screen, press
Run Test
.
2. If a 6-Position Cell Holder is installed, place the blank in position B and the sample in
position #1.
3. Press
Measure Sample
.
Parameter
Description
Delay Time
Enters the time from Test Initiation to first measurement; allows for
sample equilibration
Adv.
Interval Time
Enters the time between repeated readings
Measure Blank
(as function key)
Selects the frequency of zeroing the instrument
Once or Every Reading
Note
If Cell Correction is ON, you must run the Setup Correction application before
you can access Run Test or Measure Samples.
Note
If Auto Save Data is ON, you must enter a Data File Name before you can
access Run Test or Measure Samples.
Note
If a 6-Postion Cell Changer is installed, place the blank in position B and the sample
in cell position #1. The instrument always uses cell position #1 to scan the sample.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...