19
Concentration Measurements—Standard Curve Application
Editing a Standard Curve
Thermo Scientific
GENESYS 10S Vis User Guide
89
All absorbance measurements are removed from the screen.
Y
To reset standards
1. With the standard curve displayed, press
Edit Standards
.
2. Highlight
Reset Standards
and press
Enter
.
All standards and measurements are removed from the screen.
Y
To select a different curve fit for a standard curve
1. With the standard curve you want to edit displayed as a graph, press
Change Fit
.
2. Highlight the curve fit to use for the standard curve and press
Enter
.
The instrument applies the selected curve fit to the data and displays the new fit.
Note
To change the curve fit for a standard curve, you must display the standard curve as
a graph, not as a table.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...