Cell Correction
Cell Correction
Thermo Scientific
GENESYS 10S Vis User Guide
29
If you selected Discrete nms mode, first specify the wavelengths using the following
procedures, and then Cell Correction.
Cell Correction measures the other cells against the blank and records, stores and dates
the measurements. From these measurements Cell Correction establishes the required
correction factors, which then are automatically applied during all subsequent tests (if
Cell Correction is activated).
Specifying Wavelengths for Discrete nms Mode
Y
To specify wavelengths for Discrete nms mode
1. Highlight
Sample Positioner
and press
Enter
to set this parameter to either Auto 3
(when using three large cell holders) or Auto 6 (when using six small cell holders).
2. Highlight
Number of Matched Cuvettes
and press
Enter
.
3. Specify the number of cells you are matching and press
Enter
.
4. Press
Set nms
to select the wavelengths for which Cell Correction will be run.
A list of wavelengths appears.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...