Thermo Scientific
GENESYS 10S Vis User Guide
27
5
Cell Correction
Every test setup screen provides access to Cell Correction.
Before running Cell Correction:
• Clean the inside and outside of all the cells to be matched.
• Fill the cells with distilled water (or other blank solution), and place them in the sample
compartment (see
“Selecting and Positioning Cuvettes”
). Place the blank
cuvette in Cell “B” of the cell changer.
Cell Correction
Y
To run Cell Correction
1. Load the test type or stored test.
2. If Cell Correction is not visible, highlight
More parameters
and press
Enter
.
Note
Cell Correction is not active from the Main (Basic ATC) screen.
Note
Cell Correction is active only when the 6-Position Cell Holder is set to either Auto
6 or Auto 3. The feature is not active when the cell holder is set to 1-Cell or Manual 6, nor
when the Single Cell Holder is installed.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...