Concentration Measurements—Standard Curve Application
Measuring Samples
86
GENESYS 10S Vis User Guide
Thermo Scientific
If a 6-Position Cell Holder is installed, press the cell position buttons to reposition the cell
holder and measure the rest of the standards manually.
When all of the standards have been measured, the Standards screen (see
) shows the absorbance of each standard, along with the
slope, intercept and correlation coefficient of the standard curve.
Using the Standards Screen
Here is an example of the Standards screen showing measurement results:
You can use this screen for:
• Displaying a graph of the standard curve data (press
View Graph
)
•
(press
Save Test
)
•
(press
Edit Standards
)
•
(press
Run Test
)
Measuring Samples
Y
To measure samples automatically using the calibration curve (using Auto 6 or Auto
3)
1. Press
Run Test
.
2. Install the blank and samples.
3. Press
Enter
.
The Standard Curve screen shows the absorbance and concentration of each sample.
To switch between tabular and graphical displays, press
View Graph/Tabular
.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...