Performance Verification
Wavelength Accuracy - Standard
102
GENESYS 10S Vis User Guide
Thermo Scientific
When running the Wavelength Accuracy test, remember:
• You need Wavelength Accuracy standards designed to measure the wavelength accuracy at
the specified wavelengths.
• Use an empty cell holder as the blank.
• Measure the standards in the order they are listed on the test screen.
• You can measure the wavelength standard at only five wavelengths.
Y
To run the Wavelength Accuracy - Standards test
Highlight
Wavelength Accuracy - Standards
and press
Enter
.
Y
To add a wavelength
1. Press
Add nm
and enter the wavelength value in the
Entry
field.
2. Press
Add nm
again to add the wavelength to the list.
3. Enter the tolerance for the entered wavelength in the
Entry
field.
4. Press
Enter
.
Y
To delete a wavelength
1. Highlight the appropriate wavelength.
2. Press
Delete nm
.
Y
To run the test
1. Verify that the wavelengths and tolerances are set correctly.
2. Press
Start Test
.
The results indicate pass or fail for each wavelength.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...