Performance Verification
Photometric Accuracy
Thermo Scientific
GENESYS 10S Vis User Guide
105
Y
To add a standard
1. Press
Add Std
and enter the wavelength value in the
Entry
field.
2. Press
Enter
or
Add nm
to add the wavelength to the list.
3. Enter the absorbance or %T value for the entered wavelength in the
Entry
field.
4. Press
Enter
.
5. Enter the tolerance for the entered wavelength in the
Entry
field.
6. Press
Enter
.
The test screen displays the values you just entered for that standard.
7. Press
Start Test
to begin the measurement or
Press Esc
to save the test.
Deleting Standards
Y
To delete a standard
1. Highlight the appropriate standard.
2. Press
Delete Std
.
Running the Test
In the Photometric Accuracy screen, make sure the wavelengths, absorbance (or %T) values
and tolerances are set correctly.
Y
To run the Photometric Accuracy test
Press
Start Test
.
The results indicate pass or fail for each wavelength.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...