21
Performance Verification
Wavelength Accuracy - Internal
Thermo Scientific
GENESYS 10S Vis User Guide
101
Wavelength Accuracy - Internal
This test locates the peaks of the internal xenon lamp and displays the expected and measured
wavelengths for the peaks.
A xenon lamp has strong, fundamental lines at 229 nm, 529 nm and 883 nm. These lines are
an essential property of xenon and serve as a fundamental standard.
When running the internal standard test, remember that:
• The wavelengths and tolerance values are preset and cannot be changed.
• The cell holder should be empty.
Y
To run the Wavelength Accuracy - Internal test
1. Highlight
Wavelength Accuracy - Internal
and press
Enter
.
2. Press
Start Test
.
The results indicate pass or fail for each wavelength.
If the test fails, follow these guidelines:
• Repeat the test twice to verify that the test is failing consistently.
• Contact technical support for more troubleshooting advice.
Wavelength Accuracy - Standard
This test measures the absorbance of a wavelength accuracy standard and compares the
location of the peaks with precisely known values at up to five wavelengths. By default,
wavelength accuracy tests are performed in absorbance mode; however, this test may also be
performed in %T mode. Typical wavelengths and tolerances are included in the firmware, but
these values can be changed to match the calibration certificate included with your standards.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...