Servicing
Fan Replacement
Thermo Fisher Scientific
Model 49
i
Instruction Manual
7-11
4. Loosen the four captive screws on the mounting plate and remove the
pump.
5. Install the new pump by following the previous steps in reverse.
Fan Replacement
Use the following procedure to replace the fan (
Equipment Required:
Fan
Philips screwdriver
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the fan guard from the fan and remove the filter.
3. Pull the power connector off the fan.
4. Remove the four fan mounting screws and remove the fan.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.