Servicing
Photometer Lamp Replacement
Thermo Fisher Scientific
Model 49
i
Instruction Manual
7-27
The Calibrate Bench Temp screen appears.
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
4. Wait at least 10 seconds for the reading to stabilize, use
and
to enter the known temperature, and press
to
save the temperature value.
5. Remove thermometer or replace thermistor if removed.
6. Install the cover.
Photometer Lamp
Replacement
Use the following procedure to replace the photometer lamp (
Equipment required:
Photometer lamp
Allen wrench, 7/64-inch and 3/32-inch
Philips screwdriver
Use the following procedure to replace the photometer lamp.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the photometer lamp cable from the LAMP connector on
the photometer board.
3. Remove the two screws from the photometer lamp access hole cover on
the rear panel and remove the cover.
4. Slide the insulation off the end of the lamp.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...