Servicing
Photometer Board Replacement
Thermo Fisher Scientific
Model 49
i
Instruction Manual
7-29
1. From the Main Menu, press
to scroll to Service > press
>
to scroll to
Lamp Setting
> and press
.
The Bench Lamp Setting screen appears.
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
2. At the Bench Lamp Setting screen, use
to
increment/decrement the Lamp Setting % until Cell A Int and Cell B
Int indicate approximately 100,000 Hz.
3. Press
to save the setting.
Photometer Board
Replacement
Use the following procedure to replace the photometer board (
Equipment required:
Photometer board
Philips screwdriver
Use the following procedure to replace the photometer board.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the cables from the board. Note the locations to facilitate
reconnection.
3. Remove the two screws from the top of the board.
4. Pop off the board from the bottom mounting studs and remove the
board.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...