The Flow A Calibration menu is used to calibrate the flow A sensor to zero,
span, or restore factory default values. The flow A calibration is visible only
when the instrument is in service mode. For more information on the
service mode, see “Service Mode” earlier in the chapter. The Flow B
Calibration menu functions the same way.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Flow Calibration A
or
B.
Calibrate Flow Zero
The Calibrate Flow A Zero screen calibrates the flow sensor at zero flow.
The Calibrate Flow B Zero screen functions the same way.
Note
The pump must be disconnected before performing the zero
calibration.
RESTORE DEFAULT CAL:
RESTORE DEFAULT CAL:
RESTORE
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
FLOW SENSOR CAL:
>ZERO 200
SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...