Operation
Diagnostics Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-67
Temperatures
The Temperatures screen (read only) displays the current bench
temperature, bench lamp temperature, and O
3
lamp temperature (if the
ozonator option is installed).
•
In the Main Menu, choose Diagnostics >
Temperatures
.
Pressure
The Pressure screen (read only) displays the current bench pressure. The
pressure is measured by a pressure transducer.
•
In the Main Menu, choose Diagnostics >
Pressure
.
Flows
The Flows screen (read only) displays the current flow rate through Cell A
and Cell B. These flows are measured by internal flow sensors. For more
information, see Chapter 1, “Operations”.
•
In the Main Menu, choose Diagnostics >
Flows
.
TEMPERATURES:
>BENCH 32.3
o
C
BENCH LAMP 55.2
o
C
O3 LAMP 68.8
o
C
RANGE AVG DIAGS ALARM
PRESSURE: 753.4 mmHg
RANGE AVG DIAGS ALARM