Troubleshooting
Troubleshooting Guides
Thermo Fisher Scientific
Model 49
i
Instruction Manual
6-5
If using sample particulate
filter make sure it is not
blocked. Disconnect sample
particulate filter from the
sample bulkhead, if flow
increases, replace the filter.
Alarm - Intensity A
Alarm - Intensity B
Pre-amp Gain not set properly
Check Gain adjustment.
Defective measurement interface
board
Replace measurement
interface board.
Alarm - Zero Check
(Opt)
Alarm - Span Check
(Opt)
Alarm - Zero Autocal
(Opt)
Alarm - Span Autocal
(Opt)
Instrument out of calibration
Recalibrate instrument.
Check gas supply. Perform
manual calibration.
Alarm - Oz Level 1
Check (Opt)
Oz Level 2 Check (Opt)
Oz Level 3 Check (Opt)
Oz Level 4 Check (Opt)
Oz Level 5 Check (Opt)
Alarm - O
3
Conc.
Concentration has exceeded range
limit
Check to insure range
corresponds with expected
value. If not select proper
range.
Concentration low
Check user-defined low set
point, set to zero.
Alarm - Motherboard
Status
Alarm - Interface
Status
Alarm - I/O Exp Status
Internal cables not connected
properly
Board is defective
Check that all internal cables
are connected properly.
Recycle AC power to
instrument. If still alarming,
change board.
Table 6-2.
Troubleshooting - Alarm Messages, continued
Alarm Message
Possible Cause
Action
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...